Helix sterilization test for sterilizers - 100 tests
Helix sterilization test for sterilizers - 100 tests
Shipped in 24/48 h
Couldn't load pickup availability
Description
Description
Discover the HELIX TEST STERILIZATION TEST, an essential product to ensure the safety and effectiveness of your sterilization processes. This pack contains 100 strips designed to verify steam penetration in hollow bodies, ensuring that your equipment, such as handpieces, turbines and tubes, is properly sterilized.
Product Features
- High-quality support: made of non-toxic paper with a weight of 170 g/m², it guarantees strength and reliability.
- Practical dimensions: each strip measures 5 x 75 mm, ideal for use in various applications.
- Indicator ink: water- and heat-resistant, applied by screen printing in an elegant light blue color.
- Safe composition: the ink contains no lead or heavy metals, and is composed of six components for a precise chemical reaction.
- Classification: according to EN ISO 11140-1, the indicator is classified as Class 2 (specific test) and Class 6 (emulator indicator).
- Sensitivity to sterilization parameters: designed to respond to time, temperature and saturated steam.
- Ink color change: if the sterilization cycle is correct, the ink changes color from light blue to dark green, providing a clear visual indication.
- Latex-free: also suitable for people with latex allergies or sensitivities.
Instructions for Use
- Open the HELIX cannula.
- Fold the strip along the crease, with the indicator ink facing inward or outward.
- Insert the folded strip into the cannula capsule.
- Close the HELIX cannula again.
- Perform a sterilization cycle.
- At the end of the test, open the capsule and remove the indicator.
- Check the ink color change: if the area has completely changed from light blue to dark green, it means that saturated steam has effectively penetrated the cannula.
- Use the adhesive side of the strip for convenient permanent filing.
Sterilization Parameters
To achieve the color change of the indicator ink, the following parameters must be observed:
- 134°C saturated steam: for 3.5 minutes.
- 121°C saturated steam: for 15 minutes.
Why Choose the Helix Test Sterilization Test?
Choosing the HELIX TEST STERILIZATION TEST means investing in the safety and reliability of your sterilization procedures. Compliance with current regulations, ease of use and clear results make this test an indispensable ally for medical and dental professionals. Make sure your equipment is always ready and safe to use by choosing a high-quality product like our Helix Test.
Technical specifications
Technical specifications
| Material | Carta |
| Safety Features | Latex free |
| Product Type | Test sterilizzazione |
| Composition | carta atossica; inchiostro a sei componenti |