Reusable MIR Turbine
Description:
Reusable turbine for long-term use:
- Accurate in measurement
- Reliable over time
- Easy to clean
Does not require calibration as it is pre-calibrated at the factory
As requested by the manufacturer, please provide, by entering them in the order notes, the model and serial number of the spirometer to which the turbine will be associated.
How to Clean and Disinfect the Turbine:
Absolutely avoid placing the flow sensor under running water. The procedure must be carried out using the “immersion” method.
Procedure:
Remove the turbine from the spirometer and visually check for foreign bodies.
Immerse the turbine in the cold solution, shake it and leave it to soak for no more than 15 minutes.
Rinse the sensor in water without exposing it to the tap jet.
Place the turbine on a clean surface in an upright position and let it dry.
Cleaning Tips:
Use cold solutions based on oleyl-dimethylbenzyl-ammonium chloride or equivalent. The concentration must not exceed 15%-20%.
Technical specifications
Technical specifications
| Product Type | Turbina |